Trigger[1]|2:Source Bus|External|Hold|Immediate|Internal[[1]|2 - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

13
TRIGger Subsystem
Parameters
Item
source Available trigger sources:
The trigger source is set to IMMediate on instrument power-up and when
NOTE
entering local mode.
The MEASure and CONFigure commands automatically set the trigger source to
IMMediate.
The READ? or MEASure commands should not be used if the trigger source is set
to BUS or HOLD.
Example
Reset condition
On reset, the trigger source is set to IMMediate.
544
Description/Defaul t
– BUS: the trigger source is the group execute trigger <GET> bus
command, a *TRG common command or the
TRIGGER:IMMediate SCPI command.
– EXTernal: the trigger source is the trigger input in the back panel.
– HOLD: triggering is suspended. The only way to trigger the power
meter is to use TRIGger:IMMediate.
– IMMediate: the trigger system is always true. If
INITiate:CONTinuous is ON the power meter is continually
triggering free (free run mode). If an INITiate:IMMediate
command is sent a measurement is triggered then the power meter
returns to the idle state.
– INTernal: either INT1 (channel A) or INT2 (channel B).
TRIG:SOUR IMM
This command configures channel A for
immediate triggering.
Keysight EPM-P Series Programming Guide
Range of values
BUS
EXTernal
HOLD
IMMediate
INTernal[[1]|2]