Service:sensor[1]|2:Pcalfactor <Cal_Factor_Data - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SERVice:SENSor[1]|2:PCALfactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor data
from, the currently connected sensor. The whole calibration factor block must be
written at once as a checksum is generated. The new block must not be larger
than the existing block.
This command applies to E9320 Series sensors for peak path data only. For E4410
Series, E9300 Series, and E9320 Series sensors, average path data, refer to
SERVice:SENSor[1]|2:CALFactor <cal_factor_data>.
Syntax
Parameters
Item
cal_factor_data
Query
SERVice:SENSor[1]|2:PCALfactor?
The query returns the current peak path calibration factor block.
Query example
Error messages
– If no power sensor is connected, error -241 "Hardware missing" occurs.
– If a non E9320 Series sensor is connected, error -241
– If INIT:CONT is not set to OFF, error -221, "Settings conflict" occurs.
Keysight EPM-P Series Programming Guide
SERV
:SENS
SERV:SENS:PCALf?
"Hardware missing" occurs.
Space
1
:PCALf
2
Description/Defaul t
A binary data block. Refer to Appendix A for further information.
This command returns the peak path
calibration factor block for channel A.
SERVice Subsystem
cal_factor_data
?
15