System Tests and Adjustments
Test Management Environment (TME) Performance Tests (N5290/1 A Millimeter
System Only)
Test Management Environment (TME) Performance Tests (N5290/1
A Millimeter System Only)
The performance tests verify the electrical performance of your N5290/1 A
Millimeter System. For more information, refer to the TME webpage at
http://cal.software.keysight.com
http://cal.software.keysight.com/PNA/Help/N7840AWebHelp.htm.
This section lists the test equipment needed to run performance tests. Models
for each test equipment category are given in order of preference. Alternate
models are also listed but should be considered a lower priority. All equipment
is Keysight, Agilent, or Hewlett Packard unless otherwise specified. Refer to
each individual test for specific setup configurations. The computer and all test
equipment are connected to the unit under test (UUT) using GPIB.
Requires a PC controller to run the (TME) Performance Tests.
For a list of the equipment required for your system, refer to the N7840A
Calibration Application Help for PNA-X Series Network Analyzer Help and
to the "Test Equipment" topic.
http://cal.software.keysight.com/PNA/Help/N7840AWebHelp.htm

Guidelines for Ordering Power-Sensor Calibration Service

Power sensor calibrations services:
"Keysight Cal + Measurement Uncertainty + GB" that follows ISO
17025/ILAC-G8 standards.
(http://cal.software.keysight.com/PNA/Help/N7840AWebHelp.htm)
Use the following guidelines to ensure power sensors used by the N7800A
application receive the correct calibration service:
1. Order Option 1A7 when ordering new power sensors that will be used as
2. Order the Keysight Cal+ Measurement Uncertainty + GB for power sensor
3. When ordering periodic calibration for instruments used as lab standards
Keysight N5290/1A Service Guide
working standards in N7800A TME calibrations.
recalibration.
in the N7800A software, we recommend using "Keysight calibration +
uncertainties" for power sensors, and "Keysight calibration + uncertainties
+ guardbanding" for all other items (please visit Selecting the Right
Calibration Services). The N7800A software incorporates the ISO GUM
Uncertainty in point-to-point uncertainty calculations. The special
"H-series" calibration options in this table provide lower measurement
uncertainties through use of direct comparison to devices directly