[SENSe[1]]|SENSe2:TRACe:UNIT <character_data>
This command sets the units for the trace for the specified channel.
Syntax
Parameters
Item
character_data
Example
Reset condition
On reset the units are set to dBm.
Query
[SENSe[1]]|SENSe2:TRACe:UNIT?
The query command returns the current value of character_data.
Keysight EPM-P Series Programming Guide
SENS
1
2
SENS2:TRAC:UNIT W
:UNIT
:TRAC
Description/Defaul t
– DBM: dBm.
– W: Watts.
SENSe Subsystem
Space
?
Range of values
DBM
W
This command sets the trace units for
channel B to Watts.