The iDEN set-up provides:
– Average power in one iDEN training and data pulse.
– Peak-to-average one iDEN training and data pulse.
– Average power in a 90ms iDEN frame.
The measurement is started by detecting the iDEN training burst—for example the
burst emitted by a mobile—using the internal RF level trigger. Time gating is used
to measure the average power in the following 15ms (data pulse). Gate 1 is used
to measure this data pulse. The 90ms frame is also captured to measure the
average power in the entire frame. Gate 2 is used to measure the 90ms frame.
The following table shows the power meter presets when <character_data> is
set to IDEN. Commands not listed are preset according to their DEFault values
(for further information refer to Table 11-1):
Table 11-23
Command
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX] +800.000 MHz
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
Sensor video band wid th setup
[SENS[1]]|SENS2:BAND|BWID:VID
Gate setup
[SENS[1]]|SENS2: SWE[1]|2|3|4:
OFF:TIME
[SENS[1]]|SENS2: SWE[1]|2|3|4:
TIME
Trigger setup
Keysight EPM-P Series Programming Guide
iDEN: Power meter presets
SYSTem Subsystem
Setting
Comments
Frequency setting
NORM
Measurement mode
E9321A/25A: OFF
Sensor video bandwidth
E9322A/26A: OFF
E9323A/27A: OFF
Gate 1: 10 us
Delay between trigger point and
Gates 2 - 4: 0
time gated period.
Gate 1: 15 ms
Length of time gated period for
Gate 2: 90 ms
time gated measurements.
Gates 3 - 4: 0