[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect] <string>
This command enters the name of the sensor calibration table or frequency
dependent offset table which is to be used. The CSET1 command selects the
sensor calibration table and the CSET2 command selects the frequency
dependent offset table. The calibration factor is interpolated from the table using
the setting for [SENSe[1]]|SENSe2:FREQuency.
If [SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe is set to OFF, the
NOTE
selected sensor calibration table or frequency offset table is not being used.
Syntax
1
Parameters
Item
string
Example
Reset condition
374
:
CORR
Description/Defaul t
String data representing a sensor
calibration table, or frequency dependent
offset table name.
CORR:CSET1 "PW1"
:CSET
1
:CSET2
This command enters the name of the
sensor calibration table which is to be used
on channel A.
Keysight EPM-P Series Programming Guide
:SEL
Space
?
Range of values
Any existing table name
(Existing table names
can be listed using
MEMory:CATalog:TABle?).
string