Sense[1]]|Sense2:Sweep[1]|2|3|4 Commands - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

9
SENSe Subsystem

SENSe[1]]|SENSe2:SWEep[1]|2|3|4 Commands

These commands set offset time and time gate length as illustrated in the
following diagram:
Offset time and time gate length values can be set for up to four measurement
gates per channel. Measurement gate number is defined by the numeric value
following the SWEep component of the command.
These commands can only be used with E9320 sensors which must be set to
NOTE
NORMal mode.
404
Pre-
Trigger
With no
trace:
internally
calculated
time
Incoming signal
from sensor
Trigger
Point
Defined using:
TRIG:LEVel
TRIG:SLOPe
TRIG:HYSTeresis
With no trace: internally calculated time
Trigger
Delay
TRIG:DEL
Delayed
Trigger
SENSe:SWEep:OFFSet:TIME
Point
Data Collection Time
Keysight EPM-P Series Programming Guide
Post-Trigger
Time gate length:
SENSe:SWEep:TIME
Offset time: