Using the operation complete commands

The *OPC? and *OPC commands allow you to maintain synchronization between
the computer and the power meter. The *OPC? query command places an ASCII
character 1 into the power meter's output queue when all pending power meter
commands are complete. If your program reads this response before continuing
program execution, you can ensure synchronization between one or more
instruments and the computer.
The *OPC command sets bit 0 (Operation Complete) in the Standard Event Status
Register when all pending power meter operations are complete. By enabling this
bit to be reflected in the Status Register, you can ensure synchronization using the
GPIB serial poll.
Procedure
– Send a device clear message to clear the power meter's output buffer.
– Clear the event registers with the *CLS (clear status) command.
– Enable operation complete using the *ESE 1 command (standard event
– Send the *OPC? (operation complete query) command and enter the result to
– Send your programming command string, and place the *OPC (operation
– Use a serial poll to check to see when bit 5 (standard event) is set in the status
Keysight EPM-P Series Programming Guide
register).
assure synchronization.
complete) command as the last command.
byte summary register. You could also configure the power meter for an SRQ
interrupt by sending *SRE 32 (status byte enable register, bit 5).