13
TRIGger Subsystem

ABORt[1]|2]

This command removes the specified channel from the wait for trigger state and
places it in the idle state. It does not affect any other settings of the trigger
system. When the INITiate command is sent, the trigger system responds as it
did before ABORt was executed.
If INITiate:CONTinuous is ON, then after ABORt the specified channel
immediately goes into the wait for trigger state.
Syntax
Example
528
ABOR
1
2
ABOR
This command places channel A in the idle
state.
Keysight EPM-P Series Programming Guide