[Sense[1]]|Sense2:Correction:dcycle|Gain3:State <Boolean - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

9
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <boolean>
This command is used to enable and disable the pulse power measurement
feature.
The [SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 command should be
used to enter the duty cycle of the signal you want to measure.
Syntax
SENS
SENS2
Example
Reset condition
On reset, the pulse power measurement feature is disabled.
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe?
The query enters a 1 or 0 into the output buffer indicating the status of the pulse
power measurement feature.
– 1 is returned when the pulse power measurement feature is enabled.
– 0 is returned when the pulse power measurement feature is disabled.
Query example
382
:
1
CORR
CORR:DCYC:STAT 1
CORR:GAIN3:STAT?
:DCYC
:STAT
:GAIN3
This command enables the pulse
measurement feature on channel A.
This command queries whether the pulse
measurement feature is on or off.