Getting the Best Speed Performance

This section discusses the factors that influence the speed of operation (number
of readings/sec) of an EPM-P Series power meter.
The following factors are those which have the greatest effect upon measurement
speed (in no particular order):
– The selected measurement rate, i.e. NORMal, DOUBle, FAST.
– The sensor being used.
– The trigger mode (for example, free run, trigger with delay etc.).
– The output format: ASCii or REAL.
– The units used for the measurement.
– The command used to take a measurement.
In addition, in FAST mode there are other influences which are described in
mode"
The following paragraphs give a brief description of the above factors and how
they are controlled from SCPI.

Measurement rate

There are three possible speed settings NORMal, DOUBle and FAST. These are set
using the SENSe:MRATe command and can be applied to each channel
independently (E4417A only).
In NORMal and DOUBle modes, full instrument functionality is available and these
settings can be used with all sensors. FAST mode is available only for E-Series
sensors and averaging, limits and ratio/difference math functions are disabled.
Refer to "Specifications" in the EPM-P Series Power Meters User's Guide to see
the influence of these speed settings on the accuracy and noise performance of
the power meter.
Keysight EPM-P Series Programming Guide
on page 94.