Item
resolution
source list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command
[b] When the measurement result is linear this parameter represents the number of significant digits. When the
[c] E4417A only.
Example
READ2:DIFF?
Error messages
– INITiate:CONTinuous must be set to OFF, otherwise error -213, "INIT
– If TRIGger:SOURce is set to BUS or HOLD, error -214, "Trigger deadlock"
– If the resolution parameter is not the same as the current resolution setting on
Keysight EPM-P Series Programming Guide
Description/Defaul t
A numeric value for the resolution. If it is unspecified
the current resolution setting is used. If a value is
entered it should correspond to the current resolution
setting otherwise an error occurs.
This channel list specifies the channels used to
calculate the difference.
If unspecified and the current window setup is a
difference measurement then this difference setup is
used, otherwise it defaults to channel A-B (E4417A) or
A-A (E4416A).
sub-systems. The parameters must be entered in the specified order. If parameters are omitted, they will default
from the right. The parameter DEFault is used as a place holder. Specifying DEF leaves the parameter value
unchanged.
measurement result is logarithmic 1 to 4 represents a resolution of 1, 0.1, 0.01 and 0.001 respectively.
ignored" occurs.
occurs.
the specified window, error -221, "Settings conflict" occurs.
Measurement Commands
This command queries the difference
measurement on the lower window.
2
Range of values
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF
[c]
(@1),(@2)
[c]
(@2),(@1)
(@1),(@1)
[c]
(@2),(@2)
187