Memory:table:gain[:Magnitude] <Numeric_Value>{,<Numeric_Value - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

7
MEMory Subsystem
MEMory:TABLe:GAIN[:MAGNitude]
<numeric_value>{,<numeric_value>}
This command is used to enter calibration factors into the sensor calibration table
(8480 Series power sensors and N8480 Series sensors with Option CFT only) or
offsets into the frequency dependent offset table, currently selected using
MEMory:TABLe:SELect. Any previous calibration factor list, or offset list is
cleared before the new calibration factors/offsets are stored.
A maximum of 81 parameters for sensor calibration tables and 80 parameters for
frequency dependent offset tables can be sent with this command. For sensor
calibration tables only, the first parameter is the reference calibration factor, each
subsequent parameter is a calibration factor point in the sensor calibration table.
Entries in the frequency lists correspond as shown with entries in the calibration/
offset factor lists.
For sensor calibration tables the number of frequency points must be one less
than the number of calibration factor data points. This is verified when the sensor
calibration table is selected using
SENSe:CORRection:CSET1:SELect <string>.
Syntax
326
Calibration factor/Offset
Reference calibration factor
(sensor cal table only)
Calibration Factor/Offset 1
Calibration Factor/Offset 80
:TABL
MEM
"
:GAIN
:MAGN
Keysight EPM-P Series Programming Guide
Frequency
-
Frequency 1
"
Frequency 80
,
Space
numeric_value
?