10
STATus Subsystem

Status Register Set Commands

This section describes the five status register set commands. Each can be used to
examine all of the eleven status registers listed on the previous page.
To apply a command to a specific register, prefix the command with the name of
the appropriate register. For example, to apply the :ENABle command to the
STATus:QUEStionable register, use the following command:
STATus:QUEStionable:ENABle.
The Status Register Set commands detailed in this section are:
Keyword

:CONDition?

:ENABle
[:EVENt?]
:NTRansition
:PTRansition
:CONDition?
This query returns a 16 bit decimal-weighted number representing the bits set in
the Condition Register of the SCPI Register Set you require to control. The format
of the return is <NR1> in the range of 0 to 32767 (2
Condition Register remain unchanged after it is read.
Syntax
426
Parameter form
<NRf>|<non-decimal numeric>
<NRf>|<non-decimal numeric>
<NRf>|<non-decimal numeric>
:COND
?
Notes
[query only]
[query only]
15
-1). The contents of the