[Sense[1]]|Sense2:Correction:gain2[:Input][:Magnitude] <Numeric_Value - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

9
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]
<numeric_value>
This command is used to enter a channel offset value for the power meter setup,
for example cable loss. The power meter then corrects every measurement by this
factor to compensate for the gain/loss.
Entering a value for GAIN2 using this command automatically turns the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command to ON.
Syntax
Parameters
Item
numeric_value
388
:
SENS
1
SENS2
Description/Defaul t
A numeric value:
– DEF: the default is 0.00 dB.
– MIN: -100 dB.
– MAX: +100 dB.
CORR
:GAIN2
Keysight EPM-P Series Programming Guide
:INP
:MAGN
Space
numeric_value
DEF
MIN
MAX
Space
?
Range of values
-100 to +100 dB
DEF
MIN
MAX
MIN
MAX