Keysight Technologies EPM-P Series Programming Manual page 375

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

On reset the selected table is not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query example
Error messages
– If <string> is not valid, error -224, "Illegal parameter value" occurs.
– If a table called <string> does not exist, error -256, "File name not found"
– When a sensor calibration table is selected, the power meter verifies that the
– If the CSET1 command is used when an E-Series power sensor or N8480
Keysight EPM-P Series Programming Guide
CORR:CSET1?
occurs.
number of calibration points defined is one more than the number of frequency
points defined. When a frequency dependent offset table is selected, the
power meter verifies that the number of offset points defined is equal to the
number of frequency points defined. If this is not the case, error -226, "Lists
not the same length" occurs.
Series power sensor (excluding Option CFT) is connected the error -241,
"Hardware missing" occurs.
SENSe Subsystem
This command queries the sensor
calibration table currently used for channel
A.
9
375