Selecting A Sensor Calibration Table - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

– MEMory:TABLe:FREQuency?
– MEMory:TABLe:GAIN[:MAGNitude]:POINTs?
– MEMory:TABLe:GAIN[:MAGNitude]?
Modifying data
If you need to modify the frequency and calibration factor data stored in a sensor
calibration table you need to resend the complete data lists. There are two ways to
do this:
1 If you have retained the original data in a program, edit the program and
2 Use the query commands shown in

Selecting a sensor calibration table

After you have created the sensor calibration table, you can select it using the
following command:
[SENSe[1]]|SENSe2:CORRection:CSET1[:SELect] <string>
When the table is selected, the power meter verifies the number of calibration
factor points defined in the sensor calibration table is one parameter greater than
the number of frequency points. If this is not the case an error occurs.
To find out which sensor calibration table is currently selected, use the query:
[SENSe[1]]|SENSe2:CORRection:CSET1[:SELect]?
Keysight EPM-P Series Programming Guide
Query command which returns the frequencies stored in the sensor calibration
table (in Hz).
Query command which returns the number of calibration factor points stored
in the sensor calibration table.
Query command which returns the calibration factors stored in the sensor
calibration table. The first point returned is the reference calibration factor.
resend the data.
the data into your computer. Edit this data, then resend it.
Power Meter Remote Operation
"Reviewing table data"
1
on page 66 to enter
67