13
TRIGger Subsystem
INITiate[1]|2[:IMMediate]
This command sets the power meter in the wait for trigger state. When a trigger is
received, the measurement is taken and the result placed in the power meter
memory. If TRIGger:SOURce is set to IMMediate the measurement begins as
soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer. Refer
to
This command performs the same function as
NOTE
INITiate:[IMMediate]:SEQuence[1]|2.
Syntax
Example
Error messages
If the power meter is not in the idle state or INITiate:CONTinuous is ON, error
-213, "INIT ignored" occurs.
532
"FETCh[1]|2|3|4 Queries"
INIT
1
2
INIT2:IMM
on page 168 for further details.
:IMM
This command places channel B in the
wait for trigger state.
Keysight EPM-P Series Programming Guide