SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <boolean>
This command is used to enable and disable the use of the currently active sensor
calibration table (CSET1) or frequency dependent offset table (CSET2). When a
table has been selected and enabled, the calibration factors/offsets stored in it
can be used by specifying the required frequency using the
[SENSe[1]]|SENSe2:FREQuency command.
When the CSET1 command is set to ON, the reference calibration factor is taken
from the sensor calibration table and is used during calibration.
The CSET1 command is only applicable for 8480 Series power sensors and N8480
Series power sensors with Option CFT.
Syntax
SENS
1
SENS2
Example
Reset condition
On reset, the sensor calibration table and frequency dependent offset table are
not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a table is
enabled or disabled.
– 1 is returned when the table is enabled.
376
:
CORR
:CSET2
CORR:CSET1:STAT 1
1
:CSET
This command enables the use of the
currently active sensor calibration table for
channel A.
Keysight EPM-P Series Programming Guide
:STAT
Space
0|OFF
1|ON
?