Read[1]|2|3|4[:Scalar][:Power:ac]? [<Expected_Value>[,<Resolution>[,<Source List>]]] - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

2
Measurement Commands
READ[1]|2|3|4[:SCALar][:POWer:AC]?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to single
channel with relative mode off, aborts then initiates the specified channel,
calculates the measurement result and places the result on the bus. The result is a
power based measurement and is expressed in the units defined by
UNIT[1]|2|3|4:POWer.
INITiate:CONTinuous must be set to OFF, otherwise error -213, "INIT ignored"
NOTE
occurs. If TRIGger:SOURce is set to BUS, error -214, "Trigger deadlock" occurs.
Syntax
Parameters
Refer to
parameters in this command.
Item
expected_value
(for the expected power
level)
182
READ
1
:SCAL
2
3
Space
4
"Optional parameters"
Description/Defaul t
The expected power level parameter can be set to DEF
or a numeric value. If a value is entered it should
correspond to that set by CONFigure otherwise an
error occurs.
:POW
:AC
?
,
expected_value
DEF
on page 148 for additional details on the
Keysight EPM-P Series Programming Guide
,
resolution
source list
DEF
Range of values
sensor dependent
[a]
DEF