10
STATus Subsystem

Device Status Register Sets

The status registers contain information which give device status information. The
contents of the individual registers of these register sets may be accessed by
appending the commands listed in
The following command descriptions detail the SCPI register you require to
control but do not detail the register set commands.
The one device status register set is:
STATus:DEVice:
The following bits in these registers are used by the power meter:
Bit number
0
1
2
3
4
5
6
7-15
14
15
The Channel A and B sensor connected bits (bits 1 and 2), when queried with the
STATus:DEVice:CONDition? query are set to:
– 1, when a power sensor is connected.
– 0, when no power sensor is connected.
The Channel A and B sensor connected bits (bits 1 and 2), when queried with the
STATus:DEVice:EVENt? query indicate whether a power sensor has been
connected or disconnected depending on the state of the corresponding bits of
430
Decimal
Definition
weight
-
Not used
2
Channel A sensor connected
4
Channel B sensor connected (E4417A only)
8
Channel A sensor error
16
Channel B sensor error (E4417A only)
32
Channel A sensor Front/Rear
64
Channel B sensor Front/Rear (E4417A only)
-
Not used
16384
Front Panel key press
-
Bit 15 always 0
Status Register Set
Commands.