13
TRIGger Subsystem
TRIGger[:SEQuence]:DELay <numeric_value>
This command sets the delay between the recognition of a trigger event and the
start of a measurement.
The command is accepted for TRIGger[:SEQuence[1]] (channel A) only, for
NOTE
both single and dual channel power meters.
Syntax
Parameters
Item
numeric_value
Example
Reset condition
On reset, the trigger delay is set to 0 seconds.
546
TRIG
:SEQ
Description/Defaul t
The delay between the recognition of a trigger
event and the start of the measurement.
–
DEF:
Units are resolved to 50 ns.
TRIG:SEQ:DEL 0.001
Space
:DEL
the default value is 0 seconds.
This command sets a delay of 1 ms for
channel A.
Keysight EPM-P Series Programming Guide
numeric_value
DEF
?
Range of values
-1 to 1 seconds
DEF