Service:bist:tbase:state <Boolean - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

15
SERVice Subsystem
SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the rear panel trig out for
testing purposes.
If the command is set to:
– ON, the 10 MHz time base signal is sent to the rear panel trigger out connector.
– OFF, the 10 MHz time base signal is disabled.
Syntax
Example
Reset condition
On reset, the signal is disabled.
Query
SERVice:BIST:TBASe:STAT?
The query enters a 1 or 0 into the output buffer indicating the status of the 10
MHz time base testing.
– 1 is returned when the signal is enabled.
– 0 is returned when the signal is disabled.
582
SERV
:BIST
SERV:BIST:TBAS:STAT OFF
:TBASe
:STATe
This command disables the signal.
Keysight EPM-P Series Programming Guide
Space
0|OFF
1|ON
?