System:communicate:serial:control:rts <Boolean - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SYSTem:COMMunicate:SERial:CONTrol:RTS <boolean>
This command is used to set the serial interface (RS232 or RS422) RTS signal line
either high always (ON) or low always (OFF). In addition, the RTS signal line can be
tied to the condition of the interface receiver buffer (IBFull) resulting in the line
going high when the receiver is ready to accept data and low when the receiver
buffer is full (not ready for data). Also, with RTS in IBFull mode, the transmitter will
be inhibited when CTS is low.
Syntax
SYST
Example
Reset condition
On reset, the condition of the RTS signal line is unaffected.
Query
SYSTem:COMMunicate:SERial:CONTrol:RTS?
The query returns the current RTS signal line setting, ON, OFF or IBFull.
Query example
Keysight EPM-P Series Programming Guide
:COMM
:SER
SYST:COMM:SER:CONT:RTS ON
SYST:COMM:SER:CONT:RTS?
:CONT
:RTS
This command sets the RTS signal line
high (always).
This command queries the setting of the
RTS signal line.
SYSTem Subsystem
Space
ON
OFF
IBFull
?
11