4
CALibration Subsystem
CALibration[1]|2:ZERO:NORMal:AUTO <boolean>
This command provides a quick way of zeroing the NORMAL path of an E9320
Series sensor. The average only path is unaffected. The command cannot be used
to zero a non-E9320 Series sensor.
The command causes the power meter to perform its zeroing routine on the
specified channel when ONCE is selected. This adjusts the power meter for a zero
power reading with no power supplied to the power sensor.
The 0|OFF parameter is only required for the query response and is ignored in the
command. If 1|ON is selected, it causes the error -224, "Illegal parameter value"
to occur.
The command assumes that the power sensor is not connected to a power source.
Syntax
Example
CAL2:ZERO:NORM:AUTO ONCE
Reset condition
On reset, automatic zeroing is disabled.
Query
CALibration[1]|2:ZERO:NORMal:AUTO?
The query always returns a value of 0.
258
CAL
1
2
:ZERO
:NORM
This command causes the power meter
to perform a zeroing routine on channel
B.
Keysight EPM-P Series Programming Guide
Space
:AUTO
0|OFF
ONCE
?