Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]?
[MIN|MAX]
The query returns the current setting of the duty cycle or the values associated
with MIN and MAX.
Query example
Error messages
– If a duty cycle value is entered using
– If this command is used when an Keysight E-series power sensor is connected,
Keysight EPM-P Series Programming Guide
CORR:GAIN3?
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, "Settings Conflict"
occurs. However, the duty cycle value is set but the
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe command is not
automatically set ON.
the error -310, "System error;Dty Cyc may impair accuracy with ECP sensor"
occurs. If you are using the E4417A the error message specifies the channel.
SENSe Subsystem
This command queries the current setting
of the duty cycle for channel A.
9
381