Configure[1]|2|3|4 - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

2
Measurement Commands

CONFigure[1]|2|3|4?

This query returns the present configuration of the specified window/
measurement.
Syntax
The string returned depends on the setting of the CALCulate:MATH and
CALCulate:RELative:STATe commands.
The configuration is returned as a quoted string in the following format:
"<function> <expected_value>,<resolution>,<source list>"
CALCulate:MATH
(SENSe1)
(SENSe2)
(SENSe1)
(SENSe2)
(SENSe1 - SENSe2)
(SENSe2 - SENSe1)
(SENSe1 - SENSe2)
(SENSe2 - SENSe1)
(SENSe1 - SENSe1)
(SENSe2 - SENSe2)
(SENSe1 - SENSe1)
152
CONF
1
2
3
4
[a]
[a]
[a]
[a]
[a]
[a]
[a]
?
CALCulate:RELative:
STATe
OFF
OFF
ON
ON
OFF
OFF
ON
ON
OFF
OFF
ON
Keysight EPM-P Series Programming Guide
Function
:POW:AC
:POW:AC
:POW:AC:REL
:POW:AC:REL
:POW:AC:DIFF
:POW:AC:DIFF
:POW:AC:DIFF:REL
:POW:AC:DIFF:REL
:POW:AC:DIFF
:POW:AC:DIFF
POW:AC:DIFF:REL
<source list>
(@1)
(@2)
(@1)
(@2)
(@1),(@2)
(@2),(@1)
(@1),(@2)
(@2),(@1)
(@1),(@1)
(@2),(@2)
(@1),(@1)

Advertisement

Table of Contents
loading