MEMory:CLEar[:NAME] <character_data>
This command clears the contents of a specified sensor calibration table (8480
Series power sensors and N8480 Series sensors with Option CFT only), frequency
dependent offset table, or save/recall register.
Although the table remains, a MEMory:TABLe:FREQuency|GAIN:POINts? query
returns a 0 as there are no contents in the table.
For sensor calibration tables and frequency dependent offset tables, this
command is an alternative form of the MEMory:CLEar:TABLE command, the only
difference being the method in which the table is selected.
The contents cleared using this command are non-recoverable.
NOTE
Syntax
Parameters
Item
character_data
Example
MEM:CLE "HP8485A"
Keysight EPM-P Series Programming Guide
MEM
:CLE
Description/Defaul t
Contains an existing table name or save/recall register.
Space
:NAME
This command clears the contents of
sensor calibration table HP 8485A
MEMory Subsystem
character_data
Range of values
Any existing table name
or save/recall register.