Display[:Window[1]|2][:State] <Boolean - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

5
DISPlay Subsystem
DISPlay[:WINDow[1]|2][:STATe] <boolean>
This command enables/disables the upper or lower window (WINDow1 and
WINDow2 respectively) so that the display shows a single window only. The
displayed window is presented in expanded format, showing a single
measurement only: either the single measurement that was shown on the window,
or the currently selected measurement, if two measurements had been shown.
Syntax
Example
Reset condition
On reset, both windows are enabled.
Query
DISPlay[:WINDow[1]|2]:STATe?
The query enters a 1 or 0 into the output buffer indicating the state of the selected
window.
– 1 is returned if the window is enabled.
284
DISP
:WIND
DISP:WIND2:STAT OFF
DISP:WIND2:STAT 1
1
:STAT
2
This command disables the lower
window. The upper window in shown
in expanded format, displaying its
currently selected measurement.
This command enables the lower
window so that a dual window
display is once more provided.
Keysight EPM-P Series Programming Guide
Space
0|OFF
1|ON
?

Advertisement

Table of Contents
loading