Calibration Subsystem - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

4

CALibration Subsystem

CALibration Subsystem

The CALibration command subsystem is used to zero and calibrate the power
meter. It is also used to set the reference calibration factor for the power sensor
which is being used.
The numeric suffix of the CALibration command refers to a specific channel:
– CALibration1 represents channel A.
– CALibration2 represent channel B.
Zeroing and calibration of the power meter is recommended:
– When a 5°C change in temperature occurs.
– When you change the power sensor.
– Every 24 hours.
– Prior to measuring low level signals. For example, 10 dB above the lowest
The following CALibration commands are overlapped commands:
– CAL:ALL
– CAL:AUTO
– CAL:ZERO:AUTO
An overlapped command allows the instrument to continue parsing and executing
subsequent commands while it is still executing.
244
This command does not apply to the single channel E4416A power meter and
results in the error "Header suffix out of range."
specified power for your sensor.
Keysight EPM-P Series Programming Guide

Advertisement

Table of Contents
loading

Table of Contents