Service:sensor[1]|2:Frequency:maximum - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SERVice:SENSor[1]|2:FREQuency:MAXimum?

This query returns the maximum frequency that can be measured by the currently
connected sensor. It is applicable to E-Series sensors only. Maximum frequency
information is stored in the sensor's EEPROM.
Syntax
Example
Error messages
– If no sensor is connected, error -241, "Hardware missing" occurs.
– If a non E-Series is connected, error -241 "Hardware missing" occurs.
– If the E-Series sensor currently connected does not contain the necessary
Keysight EPM-P Series Programming Guide
SERV
:SENS
SERV:SENS2:FREQ:MAX?
information in EEPROM, error -241 "Hardware missing" occurs.
1
:FREQ
2
This query returns the maximum frequency
that can be measured by the E-Series
sensor currently connected to channel B.