STATus:OPERation:LLFail[:SUMMary]

The operation status lower limit fail summary register set contains information on
the lower limit fail status of the power meter.
The following bits in these registers are used by the power meter:
Bit number
0
1
2
3
4
5
6
7-15
The appropriate bits are set if a channel lower limit test fails or a window lower
limit test fails.
These bits are cleared if a measurement is made and the test is enabled and
passes.
Syntax
Keysight EPM-P Series Programming Guide
Decimal
Definition
weight
-
Not used
2
Channel A LLFail Status
4
Channel B LLFail Status (E4417A only)
8
Upper window LLFail Status
16
Lower widow LLFail Status
32
Upper window lower measurement LLFail Status
64
Lower window lower measurement LLFail Status
-
Not used
STAT
:OPER
:LLF
:SUMM
STATus Subsystem