2
Measurement Commands
READ[1]|2|3|4[:SCALar][:POWer:AC]:RATio:RELative?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to ratio mode
with relative mode on, aborts then initiates both channel A and B, calculates the
ratio measurement result using the new sensor data and places the result on the
bus. The result is a ratio based measurement and is expressed in the units defined
by UNIT[1]|2|3|4:POWer:RATio. The relative value used is that set by the
CALCulate:RELative:MAGNitude:AUTO command.
INITiate:CONTinuous must be set to OFF on both channels, otherwise error
NOTE
-213, "INIT ignored" occurs. If TRIGger:SOURce is set to BUS on either channel,
error -214, "Trigger deadlock" occurs.
Syntax
Parameters
Refer to
parameters in this command.
Item
expected_value
(for the expected power
level)
192
READ
1
:SCAL
2
3
Space
4
"Optional parameters"
Description/Defaul t
The power meter ignores the numeric value entered in
this parameter. Any value entered is treated like DEF.
:POW
:AC
:RAT
,
expected_value
DEF
on page 148 for additional details on the
Keysight EPM-P Series Programming Guide
:REL
?
,
resolution
source list
DEF
Range of values
sensor dependent
[a]
DEF