Setting Measurement Limits; Setting Limits - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

1
Power Meter Remote Operation

Setting Measurement Limits

You can configure the power meter to detect when a measurement is outside of a
predefined upper and/or lower limit value.
Limits are window or measurement display line based and can be applied to
power, ratio or difference measurements. In addition, the limits can be set to
output a TTL logic level at the rear panel Rmt I/O port when the predefined limits
are exceeded.

Setting limits

The power meter can be configured to verify the power being measured against
an upper and/or lower limit value. The range of values that can be set for lower
and upper limits is -150.00 dBm to +230.00 dBm. The default upper limit is
+90.00 dBm and the default lower limit is -90.00 dBm.
A typical application for this feature is shown in
Figure 1-5
84
Swept Source
OUT
Limited checking application
Device
Under Test
IN
OUT
Keysight EPM-P Series Programming Guide
Figure
1-5.
Power Meter
CHANNEL A
INPUT

Advertisement