SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut][:MAGNit
ude] <numeric_value>
This command is used to enter a gain correction value for the calibration factor.
The power meter corrects every measurement by this factor to compensate for the
gain.
Either CFACtor and GAIN1 can be used in the command—both have an identical
result. Using GAIN1 complies with the SCPI standard whereas CFACtor does not
but may make your program easier to understand.
Syntax
SENS
SENS2
Parameters
Item
numeric_value
(for CFACtor and
GAIN1)
Keysight EPM-P Series Programming Guide
:
CORR
1
Description/Defaul t
A numeric value.
– DEF: the default value is 100%.
– MIN: 1%.
– MAX: 150%.
:GAIN
1
:CFAC
Space
?
SENSe Subsystem
:INP
:MAGN
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
Range of values
[a]
1 to 150 PCT
DEF
MIN
MAX
9
371