Keysight Technologies EPM-P Series Programming Manual page 183

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Item
resolution
source list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command
[b] When the measurement result is linear this parameter represents the number of significant digits. When the
Example
READ2:POW:AC?
Error messages
– INITiate:CONTinuous must be set to OFF, otherwise error -213, "INIT
– If TRIGger:SOURce is set to BUS or HOLD, error -214, "Trigger deadlock"
– If the expected value and resolution parameters are not the same as the
Keysight EPM-P Series Programming Guide
Description/Defaul t
A numeric value for the resolution. If it is unspecified
the current resolution setting is used. If a value is
entered it should correspond to the current resolution
setting otherwise an error occurs.
The channel which the command is implemented on.
If unspecified the current window setup is used.
However, on the E4417A, if the window shows a ratio
or difference measurement, the upper window
defaults to channel A and the lower window to channel
B.
sub-systems. The parameters must be entered in the specified order. If parameters are omitted, they will default
from the right. The parameter DEFault is used as a place holder. Specifying DEF leaves the parameter value
unchanged.
measurement result is logarithmic 1 to 4 represents a resolution of 1, 0.1, 0.01 and 0.001 respectively.
ignored" occurs.
occurs.
current expected value and resolution settings on the specified window, error
-221, "Settings conflict" occurs.
Measurement Commands
This command queries the lower
window's measurement.
2
Range of values
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF
(@1)
(@2) (E4417A only)
183