15
SERVice Subsystem

SERVice:SENSor[1]|2:FREQuency:MINimum?

This query returns the minimum frequency that can be measured by the currently
connected sensor. It is applicable to E-Series sensors only. Minimum frequency
information is stored in the sensor's EEPROM.
Syntax
Example
Error messages
– If no sensor is connected, error -241, "Hardware missing" occurs.
– If a non E-Series is connected, error -241 "Hardware missing" occurs.
– If the E-Series sensor currently connected does not contain the necessary
592
SERV
:SENS
SERV:SENS1:FREQ:MIN?
information in EEPROM, error -241 "Hardware missing" occurs.
:FREQ
1
2
This query returns the minimum frequency
that can be measured by the E-Series
sensor currently connected to channel A.