OUTput Subsystem
OUTPut:TTL[1]|2:FEED <string>
This command controls which limit test drives a given TTL output. There are two
TTL outputs:
– Output 1 (TTL1).
– Output 2 (TTL2).
Both can be connected to any of the CALC subsytems (that is the upper window
upper/lower measurement or the lower window upper/lower measurement) using
OUTput:TTL:FEED <string>.
Syntax
OUTP
Parameters
TTL[1|2] String
1
344
:TTL
1
2
"CALC1|2|3|4:LIM:LOW
"CALC1|2|3|4:LIM:UPP
"CALC1|2|3|4:LIM:LOW,
CALC1|2|3|4:LIM:UPP"
Space
:FEED
?
Description
TTL Output 1 asserted indicates a lower limit fail
"
on the appropriate window/measurement.
TTL Output 1 asserted indicates an upper limit
"
fail on the appropriate window/measurement.
TTL Output 1 asserted indicates that either the
lower or upper limit failed on the appropriate
[a]
window/measurement.
Keysight EPM-P Series Programming Guide