[SENSe[1]]|SENSe2:V2P ATYPe|DTYPe

This command is used to select the type of linearity correction that will be applied
to the channel sensors being used. For most 8480 Series power sensors, the
correct (A type or D type) linearity correction table is automatically selected.
However, for the V8486A, W8486A and E8486A (with option 100) sensors the
automatic selection must be overridden and the D type (diode) correction
selected.
This command is only applicable for V8486A, W8486A and E8486A (with option
NOTE
100) sensors.
Syntax
Example
Reset condition
On reset, the linearity correction is set for A type.
Query
[SENSe[1]]|SENSe2:V2P?
The query returns the current type of linearity correction being displayed on the
screen.