Power Meter Remote Operation

Enabling the sensor calibration table system

To enable the sensor calibration table, use the following command:
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe ON
If you set [SENSe[1]]|SENSe2:CORRection:CSET1:STATe to ON and no sensor
calibration table is selected error -221, "Settings conflict" occurs.

Making the measurement

To make the power measurement, set the power meter for the frequency of the
signal you want to measure. The power meter automatically sets the calibration
factor. Use either the INITiate, FETCh? or the READ? query to initiate the
measurement as shown in the following program segments:
INITiate Example
ABORt1
CONFigure1:POWer:AC DEF,1,(@1)
SENS1:CORR:CSET1:SEL "HP8481A"
SENS1:CORR:CSET1:STAT ON
SENSe1:FREQuency 500KHZ
INITiate1:IMMediate
FETCh1?
READ? Example
ABORt1
CONFigure1:POWer:AC DEF,2,(@1)
SENS1:CORR:CSET1:SEL "HP8481A"
SENS1:CORR:CSET1:STAT ON
SENSe1:FREQuency 500KHZ
READ1?
68
Keysight EPM-P Series Programming Guide