9
SENSe Subsystem
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
This command sets the length of the time-gated period (time-gate length) for
time-gated measurements for E9320 sensors which are set to NORMal mode. To
set an E9320 sensor to NORMal mode, refer to the command
[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
on page -390.
Syntax
Parameters
Item
numeric_value
Example
Reset condition
On reset, gate 1 is set to 100μs and other gates to 1s.
Query
SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME?
The query returns the current length of the time-gated period.
408
:
SENS
1
SENS2
Description/Defaul t
The length of the time gated period in seconds.
– DEF: the default value is 100 us
Units are resolved to 1 ns.
SENS2:SWE3:TIME 0.001
: TIME
SWE
1
2
3
4
This command sets the length to
0.001 seconds.