Setting up Measurement Conditions and Display

Setting Up Condition To Make A Beep

Step 7.
Use the cursor keys to select the CC1 TH2 field. TH2 is a parameter
proportional to the contact resistance of Lp or Lc.
Step 8.
Use the softkeys or entry keys to enter the TH2 value according to
When data is entered with the entry keys, the softkeys change to units labels
(m, x1).
Softkey
INCR++
INCR+
DECR-
DECR--
Setting Up Condition To Make A Beep
The condition to make a beep sound differs depending on the beep mode, as
shown in
Table 4-5
Condition to make a beep

Setup procedure

Step 1.
Press [Meas Setup].
Step 2.
Press the LIMIT TABLE softkey.
Step 3.
Use the cursor keys to select the BEEP field.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Description
Increments the threshold value as 0, 10 m, 20 m, 50 m, 100 m, 200 m,
500 m, 1.
Increments the threshold value in steps of 10.
Decrements the threshold value in steps of 10.
Decrements the threshold value 1, 500 m, 200 m, 100 m, 50 m, 20 m,
10 m, 0.
Table
4-5.
Mode
Off
When wrong key operation is
Fail
performed.
When an error, alarm, or other
message
Pass
is outputted.
Cond ition to make a beep
Never makes a beep.
When the sorting judgment result of
the comparator is OUT_OF_BIN,
AUX_BIN, No contact or OVLD.
When the sorting judgment result of
the comparator is from BIN1 to
Table
4-4.
BIN9.