(3) Check the Serial ID (panel ID) and Version (current FPGA version),
select the target MC version, and click [FPGA UPGRADE].
The progress status is displayed at the bottom right.
After about seven minutes, the installation is completed with the information below
appearing.
(7) Click the "X" on the upper right of the screen, and exit the jig software.
018-234-06E
l Confirm the FPGA version
(1) Restart the target SE.
(2) Start the MC Manager.
(3) Make sure that the version displayed in the SE software version
upgrade window of DR Maintenance Software is the same as the
version of the installed FPGA.
1300_70J004.ai
Upgrade other software versions (CPU, GLG, MCU, PIC) as needed.
(4) Perform "Checking for Image Problems".
If an artifact is found in the image check process, perform the calibration and then
procedure in "Checking for Image Problems" again.
1300_70J005.ai
1300_70J006.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
{MU1:1.1.2_Update SE}
{IN1:13._Checking for Image Problems}
AppxIN9-3
AppxIN9-3