(6) Wrap a plastic tape around the unconnected (remaining) cable
terminals.
u NOTE u
No cable terminal may remain depending on the X-ray high voltage generator to
be connected in some cases. If no cable terminal remains, the procedure is not
necessary.
#1 Wrap: Plastic tape
#1
018-234-02E
600_700014.ai
(7) Tie the remaining cables and the X-ray shot cable(s) together, and
retain with the reusable band (supplied accessory).
#1 Retain: Reusable band(s) (supplied accessory)
<When one X-Ray shot cable is used>
<When two X-Ray shot cables are used>
DR-ID 1300 / DR-ID 1300PU Service Manual
#1
1200_700031.ai
#1
1200_700032.ai
IN2-18
IN2-18