 MC-PC EXISTENCE
Set the availability of the PC for the MC.
<Options>
- NOT EXIST [default]: Select this option when no PC is available for the MC.
- EXIST:
Select this option when the PC is available for the MC.
 A SYNCHRONOUS MC-SHUTDOWN WITH CONSOLE
Set the termination processing queue control of the MC and the Console.
<Options>
- SYNC [default]:
Select this option when no PC is available for the MC.
- ASYNC:
Select this option when the PC is available for the MC.
 LONG TIME ACCUM MODE
Set whether long time accumulation mode is to be used or not by selecting "ON" for
"AUTO DETECT" and "OFF" for "EMERGENCY MODE".
<Options>
- OFF [default]:
Not used
- ON:
Used
REFERENCE
- It takes 10 seconds to read the images by setting "ON" to "LONG TIME ACCUM
MODE" only when the specified menus are selected in the DX Console. When the
common menus are selected, the reading time is the one that set by "READ MODE".
- When setting "ON" to "LONG TIME ACCUM MODE", set 7 minutes or more to "SLEEP
START TIME" in "SE SLEEP MODE SETTING" window.
 SHOT GO OFF DELAY TIME
Set a regulatory time to secure a required irradiation time when irradiation does not
continue for a preset time due to any reason such as a delay in communication.
[default: 0]
REFERENCE
The X-ray irradiation time is often stopped by a signal from the MP due to
communication delays, which shortens the exposure time. To ensure the irradiation
time, set the SHOTGO OFF DELAY TIME value to around "the shortage of time + 10"
ms.
018-234-06E
 FTP over SSL on MC
This setting item is added in the MC V16.5. Turn ON this only for facilities that have
enabled the VA security. This setting switches the FTP protocol used in the system to
the FTP over SSL protocol.
<Options>
- OFF [default]:
- ON:
 FTP over SSL on FCR
This setting item is added in the MC V16.5. Turn ON this only for facilities that have
enabled the VA security. This setting switches the FTP protocol used in the system to
the FTP over SSL protocol.
<Options>
- OFF [default]:
- ON:
 Xair Mode
This setting item is added in the MC V16.6. Turn ON this setting only for facilities that
have enabled the Xair. This setting sets the minimum threshold of the tube current
when detecting X-ray from 10 to 5 mA, and the SmartSwitch mode appropriate for the
low output from the Xair.
<Options>
- OFF [default]:
- ON:
DR-ID 1300 / DR-ID 1300PU Service Manual
For the normal use environment, set this item.
For the VA security environment, set this item.
For the normal use environment, set this item.
For the VA security environment, set this item.
For the normal use environment, set this item.
Set only at the Xair usage site.
MU1-16