FujiFilm DR-ID 1300 Service Manual page 193

Digital radiography
Table of Contents

Advertisement

When the gain correction data is automatically generated and stored in the HDD,
"Succeeded" appears after completion.
u NOTE u
If "NG" appeared, check the following.
- Exposure condition
- Exposure field is the entire SE.
u NOTE u
If the calibration failed, restart the MC.
(4) Click [OK].
018-234-06E
 Defect Calibration
Carries out defect calibration.
- Exposure count: 5
DR-ID 1300 / DR-ID 1300PU Service Manual
MU1-28
(1)
(2)
1200_400071E.ai
MU1-28

Hide quick links: