n Procedures for Connecting the Cable
CAUTION
Measure the voltage between the cables connecting to the RDY1-RDY1 and
the RDY2-RDY2 terminals in all techniques before connecting the X-Ray shot
cables. Check to make sure that the measured voltage conforms to the coil
voltages (100/110 VAC, 110/120 VAC, 200/220 VAC and 220/240 VAC) of the relay
to be used.
If the voltage between cables differs from the coil voltage of the relay,
commission the service personnel of the X-ray equipment to check the cable
connection.
If erroneous connection is made, the machine might get damaged. Exercise
care.
CAUTION
Use the cable and the terminal as specified below when connecting the X-Ray
shot cable and the bucky cable. When crimping the terminal, use the crimp
nippers as indicated below. If the following specifications of the cable and the
terminal are not observed, the machine might get damaged.
- Cable:
AWG18 to 22 thick
- Terminal:
FV1.25-M4 (J.S.T. Connector)
- Crimp nippers: Manufactured by IZUMI; A125
018-234-02E
INSTRUCTION
u
u
Always use a reusable band (clamp) included in the supplied accessories to retain the
cable, as the band is a UL-standard component. The manufacturer and the part No. of
the reusable band are mentioned below for your reference.
Manufacturer:
Kitagawa Industries, Co. Ltd.
Name:
Reusable band
Part No.:
LWS-3S V0
NOTE
u
u
Abut the cable terminals as shown below when fastening them together.
GOOD
Cable
Screw
Terminal
Terminal block
NO GOOD