 OPE SETTING (UPRIGHT/TABLE/FREE)
OPE SETTING corresponds to Reader Type (selector) set on the DX Console.
Make the OPE SETTING and Reader Type settings coincide with each other.
DR Maintenance Software
Corresponding
u NOTES u
- If [FPD-U] is registered in any selector, always carry out the following operations.
- Connect the stand-type X-ray shot cable to the MP.
- Specify the destination of connection via the Maintenance Software in [OPE
SETTING (UPRIGHT)] and [PORT SETTING (UPRIGHT)].
- Specify whether the exposure image is to be rotated 180 degrees via the
Maintenance Software in [UPRIGHT 180 REVERSE].
- If [FPD-T] is registered in any selector, always carry out the following operations.
- Connect the bed-type X-ray shot cable to the MP.
- Specify the destination of connection via the Maintenance Software in [OPE
SETTING (TABLE)] and [PORT SETTING (TABLE)].
- Specify the direction of the patient's head via the Maintenance Software in
[TABLE DIRECTION HEAD (for DR-ID 1422SE)].
- If [FPD-C] is registered in any selector, always carry out the following operations.
- Connect the free X-ray shot cable to the MP.
- Specify the destination of connection via the Maintenance Software in [OPE
SETTING (FREE)] and [PORT SETTING (FREE)].
018-234-06E
Selector (DX Console)
1200_400063E.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
MU1-17
MU1-17