Offset Calibration
Carries out offset calibration.
No exposure is made during offset calibration.
REFERENCE
The offset calibration takes place for all of connected SE's.
018-234-06E
(1)
1200_400069E.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
(1) Click [START].
The offset correction data generation images (4 frames) are read out from the SE
(FPD).
If the offset correction data is generated and stored in the HDD, "Succeeded"is
displayed on the pop-up window.
u NOTE u
Clicking [START] during automatic offset update (approximately 30 seconds
every 10 minutes), "Error 12700 currently unavailable" is displayed. Again click
[START] approx. 30 seconds later in this case.
REFERENCE
Whether the diagnosis is successfully completed appears on the result display
area.
An error code is displayed in some cases. Refer to the Troubleshooting manual
and take appropriate countermeasures.
(2) Click [OK].
MU1-26
MU1-26