■ Calculation of the S value
The S value of the DR-ID1300 is calculated from the stitched image with three panel
units.
In the BENEO and the CR long panel, the S value is calculated from the image from
one panel and there was a restriction in the exposure position. In the DR-ID1300,
however, an appropriate S value can be calculated from the image captured at any
place on the panel unit.
BENEO long panel image
Image boundary
Image boundary
A thin part of the
object is blackened.
018-234-03E
1300_700054E.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
AppxIN4-7